ST. JUDE MEDICAL - NEUROMODULATION OCTRODE; SCS LEAD Back to Search Results
Model Number 3186
Device Problems Migration or Expulsion of Device (1395); Device Stops Intermittently (1599); Low impedance (2285)
Patient Problems Device Overstimulation of Tissue (1991); Inadequate Pain Relief (2388)
Event Date 02/20/2013
Event Type  Injury  
Event Description
Device 2 of 2.Reference mfr report: 1627487-2014-02432.It was reported the patient experienced a strong sensation when he increased his stimulation amplitude.The patient stated he felt the sensation in his leg which then moved close to his scs ipg site.A sjm representative met with the patient and a system diagnostics showed multiple lead contacts with low impedances.The representative reprogrammed the patient's scs system.On (b)(6) 2014, the patient complained his scs stimulation was intermittent.A system diagnostics show low impedances again.X-rays taken revealed the patient's scs leads migrated down around the ipg.Follow up on (b)(6) 2014 identified the patient underwent surgical intervention to revise his scs leads.The physician repositioned the leads and secured them with anchors.The patient reported receiving effective stimulation therapy postoperative.
 
Manufacturer Narrative
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd.
plano TX
Manufacturer Contact
arnie ochoa
6901 preston rd.
plano, TX 75024
9723098090
MDR Report Key3940488
MDR Text Key4577738
Report Number1627487-2014-02433
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 05/23/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/18/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/01/2015
Device Model Number3186
Device Lot Number424269
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received05/23/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/01/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SCS IPG: MODEL 3788; IMPLANT DATE:
Patient Outcome(s) Other;
Patient Age72 YR
-
-