Model Number H74939180061570
Device Problems Entrapment of Device (1212); Positioning Problem (3009)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/16/2014
Event Type  malfunction  
Event Description
It was reported that during a stenting treatment procedure, the innova 6x150x75 stent jumped and got stuck in the sheath.No patient complication have been reported.Futher information has been requested but has not been obtained.
 
Manufacturer Narrative
Device evaluated by manufacturer: the device has not been received for analysis; therefore, a failure analysis of the complaint device could not be completed.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOVA?
Type of Device
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
Manufacturer Contact
ingrid matte
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key3942294
MDR Text Key4641121
Report Number2134265-2014-04332
Device Sequence Number1
Product Code NIP
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
SIMILAR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 06/23/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/14/2015
Device Model NumberH74939180061570
Device Catalogue Number39180-06157
Device Lot Number15244503
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/23/2014
Initial Date FDA Received07/17/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/15/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-