• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES BALSTHAL 1.6MM KIRSCHNER WIRE W/TROCAR POINT 150MM; WIRE,SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES BALSTHAL 1.6MM KIRSCHNER WIRE W/TROCAR POINT 150MM; WIRE,SURGICAL Back to Search Results
Catalog Number 292.160.10
Device Problem Material Fragmentation (1261)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Additional narrative: a product development evaluation and a manufacturing evaluation were completed: the 19 k-wires were returned in a damaged state with varying amounts of scratches.(b)(4) parts showed no damage at all.(b)(4) parts showed minimal damage, which may be the result of a normal use.(b)(4) parts were heavily damaged.The colibri device mentioned in the complaint was not provided.The heavily damaged k-wires suggest that these parts were misused.Wrong power tool clamping setting, not enough manual pressure or multiple use of these single-use items are most likely.It is mentioned in the complaint that k-wires and tools of multiple companies are being used at this site.Without knowledge of which instrument was used with which k-wire a closer investigation is not possible.Further investigation regarding the manufacturing and raw material documents shows conformity to the specification as well.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes (b)(4) reports an event in (b)(6) as follows: it was reported that there was a problem with the quick coupling getting metal particles after cleaning.The product was used with both synthes and non-synthes k-wires.The non-synthes companies are also being contacted about the issue.There was no patient involvement.This is report 10 of 20 for (b)(4).
 
Manufacturer Narrative
Device was used for treatment, not diagnosis.The aware date for this report reflects the date the additional (b)(4) parts were received by synthes as part of the existing complaint received on (b)(4) 2014.Additional common device name: hty.No nonconformance reports were generated during production.Review of the device history record showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.The investigation could not be completed; no conclusion could be drawn, as the product is entering the complaint system.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.6MM KIRSCHNER WIRE W/TROCAR POINT 150MM
Type of Device
WIRE,SURGICAL
Manufacturer (Section D)
SYNTHES BALSTHAL
dornacherstrasse 20
balsthal CH47 10
SZ  CH4710
Manufacturer (Section G)
SYNTHES BALSTHAL
dornacherstrasse 20
balsthal CH47 10
SZ   CH4710
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key3942686
MDR Text Key11951503
Report Number3009450863-2014-10022
Device Sequence Number1
Product Code LRN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PPREAMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 06/27/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/17/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number292.160.10
Device Lot Number8331600
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/27/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/28/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/04/2013
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-