• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO TPS HANDPIECE CORD; DRIVER, WIRE, AND BONE DRILL, MANUAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO TPS HANDPIECE CORD; DRIVER, WIRE, AND BONE DRILL, MANUAL Back to Search Results
Catalog Number 5100004000
Device Problem Device Remains Activated (1525)
Patient Problem No Patient Involvement (2645)
Event Date 06/27/2014
Event Type  malfunction  
Event Description
It was reported that during routine maintenance conducted by a manufacturer field service technician at the user facility the tps handpiece cord was causing the handpiece to run without user activation after the trigger was released.It was reported that there was no associated procedure.No patient involvement, no delay, no medical intervention and no adverse consequences were reported with this event.
 
Manufacturer Narrative
Per visual inspection, the engineering technician found corroded pins on the cable.The device was discarded by the manufacturer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPS HANDPIECE CORD
Type of Device
DRIVER, WIRE, AND BONE DRILL, MANUAL
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
casey metzger
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key3943460
MDR Text Key4643229
Report Number0001811755-2014-02570
Device Sequence Number1
Product Code DZJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K943540
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 06/27/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number5100004000
Device Lot Number13002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/30/2014
Initial Date Manufacturer Received 06/27/2014
Initial Date FDA Received07/18/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/02/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-