SYNTHES BRANDYWINE 6.0MM TI HARD ROD 75MM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD Back to Search Results
Catalog Number 498.103
Device Problem Break (1069)
Patient Problems Failure of Implant (1924); Pain (1994); Joint Disorder (2373); Impaired Healing (2378)
Event Type  Injury  
Event Description
It was reported that the patient had a nonunion, adjacent level degenerative disease above and below the original construct and had a broken rod from a surgery that was performed on (b)(6), 2011.The rod broke at an unknown time.During the revision surgery on (b)(6), 2014, the tip of a variable axis screw iii screwdriver broke off in the screw.The patient had a l3-l5 posterior spine fusion with titanium universal spine system on (b)(6), 2011.At an unknown point in time, the patient saw the surgeon for increasing back pain.The surgeon assessed that the patient had a nonunion at l4-l5, and also had developed adjacent level degeneration at the l2-l3 and l5-s1 level that may be responsible for his pain.The broken rod was not discovered until the revision surgery on (b)(6), 2014.For the revision surgery the surgeon removed all of the universal spine system hardware and discovered the left rod was broken.The surgeon then instrumented the patient with universal spine system dual opening from l2, l3, l4, l5, and s1.He then performed a spinal decompression.He then decided to replace the right l5 universal spine system dual opening screw with a universal spine system variable axis screw iii screw because the angle of the universal spine system dual opening screw would not permit it to be connected to the rod and needed a polyaxial screw.During the insertion of the universal spine system variable axis screw iii screw the tip of the screwdriver broke off in the hex of the screw.The surgeon then placed a short rod in the screw and connected it using a nut and collar and tightened it so the poly screw would be locked.He then used the counter torque sleeve to get the screw out using the rod.The surgeon then placed a new universal spine system variable axis screw iii 9.0 x 50mm screw on.He then connected the rods successfully and then placed his iliac screws, connected them, and successfully completed the procedure.The broken rod was between left l4 and l5 screws.Left l4 screw was a 7.0 x 55 mm screw and the left l5 screw was a 7.0 x 50mm screw.There was a twenty minute delay reported during the revision surgery.No patient harm was reported during the revision surgery.This is report 7 of 21 for (b)(4).
 
Manufacturer Narrative
Date of event: unknown.Common device name: additional product codes: mni, mnh, kwp, kwq.Device history records indicate that the reported product was manufactured in accordance with all established requirements and procedures with no deviations reported.Device history records of the reported finished goods lot further indicate that the product lot underwent all required inspection operations and was found to meet all established criteria for release to warehouse, with no product nonconformities reported.Device history records of the reported product indicate that the raw material underwent all established inspection and test requirements and was found to meet all established specifications.Review of the device history records showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
6.0MM TI HARD ROD 75MM
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD
Manufacturer (Section D)
SYNTHES BRANDYWINE
1303 goshen parkway
west chester PA 19380
Manufacturer (Section G)
SYNTHES BRANDYWINE
1303 goshen parkway
west chester PA 19380
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key3946002
MDR Text Key4642777
Report Number2530088-2014-10187
Device Sequence Number1
Product Code NKB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK090605
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Health Professional
Type of Report Initial
Report Date 06/23/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number498.103
Device Lot Number6430545
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/23/2014
Initial Date FDA Received07/21/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/19/2010
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Weight109
-
-