Catalog Number UNK-ASR
Device Problem Corroded (1131)
Patient Problems Scar Tissue (2060); Discomfort (2330); Joint Swelling (2356); No Code Available (3191)
Event Date 06/16/2014
Event Type  Injury  
Event Description
Patient contacted depuy as a result of the asr recall to initiate a claim.Medical records were obtained.Medical records indicate patient was revised due to discomfort, joint effusion and elevated metal ion levels (co 370, cr 1100).Upon revision, brownish fluid, ruptured sutures posteriorly, partially intact external rotators, extensive scar tissue, and corrosion on the femoral neck from the femoral head were noted.
 
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The asr platform was voluntarily recalled from the market in august 2010, and the asr product codes are now considered inactive.Further investigation of this individual incident will not be undertaken, as there is an ongoing investigation regarding the root cause(s) and/or corrective actions.Ref.Wwcapa (b)(4).Depuy considers the investigation closed at this time.Should the product and/or additional information be received, the investigation will be re-opened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN DEPUY ASR FEMORAL HEAD
Type of Device
HIP FEMORAL HEAD
Manufacturer (Section D)
DEPUY INTL., LTD. - 8010379
st. anthony's rd
leeds LS11 8 D
UK  LS11 8 DT
Manufacturer (Section G)
DEPUY INTL., LTD. - 8010379
st. anthony's rd
leeds LS11 8 D
UK   LS11 8 DT
Manufacturer Contact
stacey trick
700 orthopaedic drive
warsaw, IN 46581
5743714554
MDR Report Key3948252
MDR Text Key21879008
Report Number1818910-2014-23877
Device Sequence Number1
Product Code KXA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation Patient
Remedial Action Recall
Type of Report Initial
Report Date 07/01/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Catalogue NumberUNK-ASR
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/01/2014
Initial Date FDA Received07/21/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-1749/1816-2011
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age81 YR
-
-