• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION OCTRODE; SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION OCTRODE; SCS LEAD Back to Search Results
Model Number 3186
Device Problem Device Operates Differently Than Expected (2913)
Patient Problems Device Overstimulation of Tissue (1991); Stenosis (2263); Inadequate Pain Relief (2388)
Event Date 05/28/2014
Event Type  Injury  
Event Description
Device 1 of 2.Reference mfr.Report #: 1627487-2014-054444.It was reported the patient is receiving unsatisfactory pain relief due to diminished stimulation.It was also reported the patient experiences overstimulation when attempting to increase stimulation, due to the patient's spinal stenosis.
 
Manufacturer Narrative
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
plano TX
Manufacturer Contact
drew johnson
6901 preston rd.
plano, TX 75024
9725264667
MDR Report Key3950277
MDR Text Key4645092
Report Number1627487-2014-05443
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 05/28/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/01/2015
Device Model Number3186
Device Lot Number4342976
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/28/2014
Initial Date FDA Received06/19/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/01/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SCS ANCHORS: MODEL 1192 (2); SCS IPG: MODEL 3788; IMPLANT DATE:; IMPLANT DATE:
Patient Outcome(s) Other;
Patient Age87 YR
-
-