Model Number 3788 |
Device Problems
Device Inoperable (1663); Use of Device Problem (1670); Communication or Transmission Problem (2896)
|
Patient Problem
No Information (3190)
|
Event Date 06/01/2013 |
Event Type
Injury
|
Event Description
|
It was reported the patient has not charged her scs ipg in approximately a year.A sjm representative met with the patient and confirmed the patient's ipg battery is depleted and the ipg is inoperable.Surgical intervention may be undertaken at later date to address the issue.
|
|
Event Description
|
Additional information identifed the patient's scs ipg was explanted and replaced with a new one.In addition, the patient reported losing a significant amount of weight, and the ipg was causing discomfort.During the procedure, the ipg was found to have flipped inside the pocket.Additional follow-up identified effective stimulation therapy was restored postoperative, and the discomfort issue is resolved.
|
|
Manufacturer Narrative
|
Correction umber: 1627487-07262012-002-r.This ipg serial number was included in field advisories.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|
|
Manufacturer Narrative
|
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|
|
Search Alerts/Recalls
|