Device evaluation summary: biocompatibility testing to iso 10993 was successfully completed on skin-contacting surfaces of the lifevest device.The therapy electrodes contain a small amount of nickel, which is a known allergen.Evaluation of belt (b)(4) has been completed.The reported problem (gel leak) was confirmed.Upon evaluation, there was gel leaking from one blister in the rear 1 therapy electrode.There was no allegation that the gel leak contributed to the allergic reaction.Biocompatibility testing to iso 10993 was also successfully completed on the gel.
|