• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF) Back to Search Results
Model Number 97714
Device Problems Improper or Incorrect Procedure or Method (2017); Low Battery (2584)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  Injury  
Event Description
It was reported that the device had several overdischarges (od) due to the patient not charging promptly.It was unknown how many ods had occurred and how many physician mode recharges had occurred.The physician elected to replace the battery to give the patient full charge capabilities.The patient was warned that it will not be replaced if od events occur in the future.The patient was doing well after replacement.
 
Manufacturer Narrative
Concomitant products: product id 97754, serial #(b)(4), product type recharger; product id 97740, serial # (b)(4), product type programmer, patient; product id 977a260, serial # (b)(4), implanted: (b)(6) 2013, product type lead; product id 977a260, serial # (b)(4), implanted: (b)(6) 2013, product type lead.
 
Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESCAN
Type of Device
STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
ceiba norte industrial park, r
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
7000 central avenue ne rcw215
minneapolis MN 55432
Manufacturer Contact
diane wolf
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263987
MDR Report Key3952695
MDR Text Key4606212
Report Number3004209178-2014-13414
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 07/01/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/14/2014
Device Model Number97714
Device Catalogue Number97714
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/01/2014
Initial Date FDA Received07/23/2014
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received07/23/2014
08/01/2014
Date Device Manufactured08/14/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age00062 YR
-
-