Catalog Number 5400300000 |
Device Problems
Overheating of Device (1437); Smoking (1585)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/26/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that during a procedure at the user facility the core impaction drill overheated and smoke came out of the tip of the device.The procedure was completed successfully using back-up equipment without a clinically significant delay; no medical intervention and no adverse consequences were reported.
|
|
Manufacturer Narrative
|
Failure analysis is in progress; additional information will be submitted once the quality investigation is completed.
|
|
Event Description
|
It was reported that during a procedure at the user facility, the core impaction drill overheated and smoke came out of the tip of the device.The procedure was completed successfully using back-up equipment without a clinically significant delay; no medical intervention and no adverse consequences were reported.
|
|
Manufacturer Narrative
|
The technician was unable to confirm the reported events but noted that the motor assembly was corroded.
|
|