Catalog Number 5400300000L |
Device Problem
Overheating of Device (1437)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/30/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that during a procedure at the user facility, the loaner core impaction drill was overheating.The procedure was completed successfully using back-up equipment without a clinically significant delay; no medical intervention and no adverse consequences were reported.
|
|
Event Description
|
It was reported that during a procedure at the user facility the loaner core impaction drill was overheating.The procedure was completed successfully using back-up equipment without a clinically significant delay; no medical intervention and no adverse consequences were reported.
|
|
Manufacturer Narrative
|
The failure for heat was confirmed through functional testing, disassembly, and visual inspection.Through visual inspection, the bearings were brinelled.The device was repaired and placed back into stryker stock.
|
|
Manufacturer Narrative
|
Failure analysis is in progress; additional information will be submitted once the quality investigation is completed.
|
|