Device 1 of 3.Ref mfr reports: 1627487-2014-01473, 01474.It was reported the pt is without left side stimulation.The pain pattern is bilateral back and legs.After further investigation, it was revealed the lead has migrated from t8 to t11.No known traumatic events have occurred.Surgical intervention was undertaken on (b)(6) 2014 in efforts to revise the leads.During procedure, the physician noted purulent drainage and puss extending from the lead and ipg sites.Subsequently, the physician opted to explant the scs system.Culture were taken, however, results are unk at this time.
|