A hemodialysis inpatient user facility reported during initiation, lots of bubbles were visible in the line.Clinic coordinator stated they primed with saline and everything looked good, did recirculation and were ready to do treatment.As soon as the blood hit the blood pump, air bubbles were visible in the line.Clinic coordinator stated that they there was no impact to the patient and treatment was continued with a new tubing set.Nurse stated there were no defects seen on the lines.They changed out the lines to a different serial number and the issue was resolved.Sample has not been returned to the manufacturer.
|