• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK IRELAND LTD ZILVER PTX DRUG-ELUTING PERIPHERAL STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK IRELAND LTD ZILVER PTX DRUG-ELUTING PERIPHERAL STENT Back to Search Results
Catalog Number ZIV6-35-125-6.0-100-PTX
Device Problems Occlusion Within Device (1423); Improper Flow or Infusion (2954)
Patient Problems Occlusion (1984); Surgical procedure, additional (2564)
Event Date 06/09/2011
Event Type  Injury  
Event Description
On (b)(6) 2014 one ziv6-35-125-6.0-100-ptx stent (this report) and one ziv6-35-125-7.0-ptx stent were placed in the right sfa by ipsilateral antegrade approach.Although dissection of right sfa occurred during the procedure, the placement could be completed successfully.The patient had a feeling of strangeness when leaving the hospital.On (b)(6) 2014 the patient visited the hospital claiming pain in his lower limb.Since thrombosis was suspected, ct imaging was conducted, but thrombosis could not be confirmed from the ct images.On (b)(6) 2014 angiography confirmed the thrombotic occlusion in ptx stents.Aspiration of the blood clots with tvac (manufactured by nipro) was performed 5 times and after the clots were sucked in, low-pressure long inflation with jackal 4x80 (manufactured by kaneka) was conducted; however, the flow did not improve.The lesion was therefore dilated with angiosculpt 5x40 (manufactured by angioscore) next, followed by angiography, which confirmed the improvement of the blood flow and thus, the procedure was ended.After the procedure, warfarin was administered to the patient.Subjective symptom of the patient disappeared.The patient is being monitored now.
 
Manufacturer Narrative
Pma/510(k)#: p100022/s001.Additional stent involved in this complaint event (ziv6-35-125-7.0-100-ptx) will be submitted in an additional report, ref # 3001845648-2014-00119.This complaint report meets the reporting criteria of an fda mdr report based on the intervention carried out due to the occurrence of thrombosis while a zilber ptx stent was indwelling.'the procedure has been revealed to have been conducted by ipsilateral antegrade approach from "right sfa." at first, the physician planned to treat the right sfa lesion only with poba.Thus, the right sfa was directly punctured and a guiding sheath was advanced from there; however, during insertion of the guiding sheath, dissection in the lesion occurred.Ptx stents placement was therefore conducted as countermeasure against it.The delivery systems of the used two ptx were advanced through the guiding sheath advanced from right sfa.' there was no zilver ptx device of lot number c908200 in stock at the time of the complaint investigation.The stent was implanted in the patient; therefore, it is not available for evaluation.With the information provided a document based investigation was carried out.Images were requested relating to this complaint but have not been received to date.The below list indicates potential risk factors that can generally contribute to the thrombosis event: patient factors - history of coagulopathy/prior thrombosis (e.G.Dvt), diabetes, especially if poorly controlled, cancer/chemotherapy, advanced age, obesity, hyperlipidemia, hypertension, smoking.Lesion factors - long lesion, small vessel diameter, sever calcification, lesion totally occluded prior to stent placement, placement for in stent restenosis.Procedure factors - residual inflow, outflow or in-segment stenosis or dissection, poor run off (i.E.Beyond trifurcation).Medication factors - inadequate procedural heparinization, inadequate loading close of antiplatelet (ticlopidine or clopidogrel), inadequate dapt prescribed, non-responder to the apt, or non-compliant with prescribe regimen.According to the information provided the patient had the following known potential risk factors for thrombosis: smoker, coronary artery disease and pulmonary disease were also confirmed.It can be noted that the patient was on the anti-platelet therapy prior to and after the procedure; therefore, medication factor is not a contributing factor to stent thrombosis in this case.Comments provided for by the physician involved in this complaint were as follows: 'i assume this thrombosis occurred due to the dissection during procedure, astriction and/or residual stenosis in the lesion (out of where stents were placed).In addition to above, considering that incomplete expansion of the stent was not observed, i do not think the cause of this event is ptx stents.' the complaint is confirmed based on customer testimony; however, as per the above feedback it can be stated that it is unlikely that stent thrombosis occurred due to zilver ptx malfunction.A definitive root cause of this stent thrombosis cannot be determined.As per instructions for use, arterial thrombosis is noted as a potential adverse event associated with the placement of this device.Prior to distribution all zilver ptx devices are subject to visual inspection and functional checks to ensure device integrity.A review of the manufacturing records for zilver ptx and zilver ptx drug eluting stent revealed no discrepancies that could have contributed to this complaint.Quality engineering will continue to monitor complaints of this nature for potential emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZILVER PTX DRUG-ELUTING PERIPHERAL STENT
Manufacturer (Section D)
COOK IRELAND LTD
limerick
EI 
Manufacturer Contact
tracy o'sullivan
61334440
MDR Report Key3971215
MDR Text Key4631302
Report Number3001845648-2014-00118
Device Sequence Number1
Product Code NIU
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P100022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/09/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/31/2015
Device Catalogue NumberZIV6-35-125-6.0-100-PTX
Device Lot NumberC908200
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date06/09/2014
Event Location Hospital
Initial Date Manufacturer Received 06/11/2014
Initial Date FDA Received07/08/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/19/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-