• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CPI - DEL CARIBE EASYTRAK 2; IMPLANTABLE LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CPI - DEL CARIBE EASYTRAK 2; IMPLANTABLE LEAD Back to Search Results
Model Number 4543
Device Problems Failure to Capture (1081); Fracture (1260); High impedance (1291); Sensing Intermittently (1558)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/14/2014
Event Type  Injury  
Event Description
Boston scientific received information that this left ventricular (lv) lead exhibited high out-of-range pacing lead impedance, loss of capture (loc) intermittent sensing after the patient fell due to bicycle accident.The lead had greater than 2000 ohms impedance when tested with pacing system analyzer (psa) and appeared to be fractured.The lead was capped and replaced.No additional adverse patient effects were reported.
 
Manufacturer Narrative
(b)(4).As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYTRAK 2
Type of Device
IMPLANTABLE LEAD
Manufacturer (Section D)
CPI - DEL CARIBE
guidant puerto rico b. v.
dorado PR
Manufacturer (Section G)
CPI - DEL CARIBE
guidant puerto rico b. v.
dorado PR
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 55112
6515824786
MDR Report Key3971329
MDR Text Key18291348
Report Number2124215-2014-10652
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 04/16/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/17/2013
Device Model Number4543
Other Device ID NumberEASYTRAK 2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 04/16/2014
Initial Date FDA Received07/31/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/17/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
4470; 4543; 0185; N119
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age61 YR
-
-