The customer reported the coulter lh 500 hematology analyzer was generating 5c abnormal ii vote out all parameters.The customer stated that the instrument did not generate hemoglobin (hgb) result.The customer noticed a burning smell was coming from the instrument after it was placed into shutdown cycle.The customer powered off and unplugged the instrument from the electrical outlet.There was no actual smoke, only an electrical burn smell.There was no death, injury, or change to patient treatment attributed to or connected with this event.
|
A beckman coulter (bec) field service engineer (fse) was dispatched to assess the instrument.The fse observed disconnected tubing from the sweep flow check valve which caused a small amount of diluent to leak onto the peltier module (cause of burning smell), adjacent solenoid valves and the differential mixing motor.The leak was contained within the instrument.When the instrument was powered on, the instrument generated voltage errors/messages and was inoperable.The fse replaced the tubing and check valve for the sweep flow and replaced the peltier module, the solenoid valves and the differential mixing motor to resolve this issue.The fse verified the instrument and the instrument was returned to normal operation.(b)(4).
|