• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION CHARGING SYSTEM; SCS CHARGER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION CHARGING SYSTEM; SCS CHARGER Back to Search Results
Model Number 3711
Device Problems Device Operates Differently Than Expected (2913); Temperature Problem (3022)
Patient Problems Nausea (1970); Burning Sensation (2146)
Event Date 06/12/2014
Event Type  Injury  
Event Description
Device 2 of 2.Reference mfr report 1627487-2014-05477.
 
Manufacturer Narrative
This charger model was associated with a field correction.Manufacturer's evaluation: corrective and preventive action (capa) investigation was performed.Result - pocket heating confirmed.The investigation for capa (b)(4) associated with heating while charging (pocket heating) concluded that the charger was capable of transferring energy to the ipg at a rate that would cause heating of the ipg and/o charging wand of sufficient elevated temperature to cause pain and burns.The heating while charging was determined to be exacerbated by off-axis charging of shallow implanted ipgs and that all chargers were capable of elevated heating.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHARGING SYSTEM
Type of Device
SCS CHARGER
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
plano TX
Manufacturer Contact
drew johnson
6901 preston rd.
plano, TX 75024
9725264667
MDR Report Key3975362
MDR Text Key17261364
Report Number1627487-2014-05478
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeFI
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Not Applicable
Remedial Action Other
Type of Report Initial
Report Date 06/12/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/09/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/01/2014
Device Model Number3711
Device Lot Number4112517
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received06/12/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/01/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487-07262012-001-C
Patient Sequence Number1
Patient Outcome(s) Other;
-
-