LIFE TECHNOLOGIES HOLDINGS PTE. LTD. APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR; INSTRUMENTATION FOR CLINICAL MULTIP Back to Search Results
Catalog Number 4407205
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Information (3190)
Event Date 03/17/2014
Event Type  malfunction  
Event Description
A customer reported there was apparent failed amplifications of false negative in well e2.The instrument involved was applied biosystems 7500 fast dx (cat.No.4407205, serial no.(b)(4).No pt involvement reported.(b)(4).
 
Manufacturer Narrative
Device is an instrument and is not implanted/explanted.The instrument was investigated on site.Problem solved by replacing the heated cover.Device intended use: the applied biosystems 7500 fast dx real-time pcr instrument with the sds software version 1.4 is an real-time nucleic acid amplification and detection system that measures nucleic acid signals from reverse transcribed rna and converts them to comparative quantitative readouts using fluorescent detection of dual-labeled hydrolysis probes.The 7500 fast dx real-time pcr instrument is to be used only by technologists trained in laboratory techniques, procedures, and on use of the analyzer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR
Type of Device
INSTRUMENTATION FOR CLINICAL MULTIP
Manufacturer (Section D)
LIFE TECHNOLOGIES HOLDINGS PTE. LTD.
singapore
SN 
Manufacturer (Section G)
LIFE TECHNOLOGIES HOLDINGS PTE. LTD.
Manufacturer Contact
kelli tanzella
3175 staley road
grand island, NY 14072
7167743122
MDR Report Key3978848
MDR Text Key21806559
Report Number3003673482-2014-00033
Device Sequence Number1
Product Code NSU
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082562
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 03/17/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number4407205
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/17/2014
Initial Date FDA Received06/27/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-