Patient noticed air entering the system at the connector entering the better bladder (small amount) and felt it was due to the tyband connector being to close to the edge of the connector.They resolved the issue by applying another tyband further onto the connector.The territory manager was told there was no patient effects as all of the air was entrapped in the better bladder and the bubble sensor never alarmed during this situation.The territory manager also tried to determine why this was not noticed during priming and if the negative pressure was abnormally high.
|