Device 2 of 3.Reference mfr report: 1627487-2014-21497.Reference mfr report: 1627487-2014-21499.It was reported the pt ((b)(6)) was receiving only unintended left leg stimulation post-operatively.X-rays confirmed the lead had migrated.Subsequently, the pt underwent surgical intervention.Pre-operative diagnostics revealed one invalid lead contact.As a result, the lead was explanted and replaced.It was also reported the physician encountered difficulty steering the new lead (device 2) to the desired location; however, after several repositioning attempts, the lead was implanted successfully.Follow-up identified the pt was receiving stimulation in the desired pain pattern; however, the amplitude was uncomfortable for the pt.Diagnostics revealed low lead impedance contacts (including ipg as device 3).There is no further info at this time.
|