• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. GYNECARE TVT OBTURATOR SYSTEM; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. GYNECARE TVT OBTURATOR SYSTEM; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC Back to Search Results
Catalog Number 810081L
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/05/2014
Event Type  malfunction  
Manufacturer Narrative
Conclusion: the actual device involved in this event was returned for evaluation.The device was visually evaluated.Upon evaluation, one plastic tube was broken at the level of the guide extremity.The location of the break on the needle indicated it could not have happened if the needle was properly locked on the guide as it is when the product is packaged at the manufacturing site.This suggests that the needle was removed from the bulge that holds the needle at the guide.In addition, a review of the batch manufacturing records was conducted and the batch met all finished goods release criteria.
 
Event Description
It was reported that a patient underwent an obturator sling procedure on (b)(6) 2014.During the procedure, the extremity of the mesh got from the needle with the collar.The mesh was broken at the tip of the needle.Another like device was used to complete the procedure with no adverse patient consequences.Additional information has been requested.
 
Manufacturer Narrative
(b)(4).Conclusion: to date, the device has not been returned.If the product is returned for evaluation, any further info derived from the evaluation will be submitted in a supplemental 3500a form.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GYNECARE TVT OBTURATOR SYSTEM
Type of Device
MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 015
Manufacturer (Section G)
ETHICON SARL-NEUCHATEL
puits-godet 20 2000 neuchatel
neuchatel
SZ  
Manufacturer Contact
guillermo villa
route 22 west po box 151
somerville, NJ 08876
9082180707
MDR Report Key3984753
MDR Text Key12678589
Report Number2210968-2014-10890
Device Sequence Number1
Product Code OTN
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K033568
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/21/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/06/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2014
Device Catalogue Number810081L
Device Lot Number3753194
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/27/2014
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received12/01/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/16/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-