Model Number 3688
Device Problems Premature End-of-Life Indicator (1480); Device Displays Incorrect Message (2591); Unintended Movement (3026)
Patient Problems Fall (1848); Inadequate Pain Relief (2388)
Event Date 07/09/2014
Event Type  Injury  
Event Description
It was reported the patient is without stimulation and received a low battery warning.The patient suffered a fall in the past and the ipg feels as if it might be slanted.The sjm representative met with the patient but troubleshooting was unable to resolve the issue.A replacement programmer was also unable to resolve the issue.Program parameters indicate premature battery depletion.Surgical intervention may take place at a later date to address the issue.
 
Manufacturer Narrative
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EONC
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
plano TX
Manufacturer Contact
tanja hoffman
6901 preston rd.
plano, TX 75024
9723098520
MDR Report Key3986463
MDR Text Key20108819
Report Number1627487-2014-26606
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 07/09/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/31/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/01/2016
Device Model Number3688
Device Lot Number4424723
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received07/09/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
IMPLANT:; SCS LEAD: MODEL 3156; IMPLANT:; SCS EXTENSION: MODEL 3386; SCS EXTENSION: MODEL 3386; IMPLANT:
Patient Outcome(s) Other;
Patient Age67 YR
-
-