• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SINGAPORE OPERATIONS ATTAIN ABILITY PLUS; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SINGAPORE OPERATIONS ATTAIN ABILITY PLUS; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE Back to Search Results
Model Number 429688
Device Problems Device Contamination with Body Fluid (2317); High Capture Threshold (3266)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/30/2014
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the full lead was returned and analyzed.Analysis was performed and no anomalies were found.The outer insulation of the lead was observed to have blood ingression.Visual summary analysis of the lead indicated damage at implant.
 
Event Description
It was reported that during an implant attempt, the left ventricular (lv) lead was positioned in the coronary sinus and had high thresholds.The physician noted there was quite a lot of blood inside the lead when it was extracted from the body.Another lead was implanted.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
This event occurred outside the us where a similar model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.(b)(4).
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN ABILITY PLUS
Type of Device
DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE
Manufacturer (Section D)
MEDTRONIC SINGAPORE OPERATIONS
nasaco tech centre, 49 changi
singapore
SG 
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM DISEASE MGMT
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
nashoane fulwood-kelley
8200 coral sea st ne
mounds view, MN 55112
7635260583
MDR Report Key3990136
MDR Text Key12267822
Report Number3008973940-2014-00288
Device Sequence Number1
Product Code OJX
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P080006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,company representati
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 06/30/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/07/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/13/2015
Device Model Number429688
Device Catalogue Number429688
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/03/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/26/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/02/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age00079 YR
Patient Weight69
-
-