IPG MFG SWITZERLAND PROTECTA XT CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
|
Back to Search Results |
|
Model Number D354TRG |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Syncope (1610); Dyspnea (1816); Fainting (1847); Weakness (2145)
|
Event Date 05/01/2014 |
Event Type
Injury
|
Event Description
|
It was reported that the patient experienced shortness of breath and weakness, and collapsed after sporting activity.It was noted that the patient did not experience typical syncope, and the ecg (electrocardiogram) was normal.A few days later, the lower rate was reprogrammed from 40 to 60 bpm, and mode changed from ddd to dddr.The event was noted to be related to the programming.The device remains in use.No further patient complications have been reported as a result of this event.The device was noted to be part of the prompt study.
|
|
Manufacturer Narrative
|
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.This event occurred outside the us where the same model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.This model number is not approved for distribution in the united states, however, it is similar to a device marketed in the u.S.The event is being reported due to an alleged malfunction.Products: 4196 lead; 5076 lead.(b)(4).
|
|