Model Number 5392
Device Problem Power Problem (3010)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/24/2014
Event Type  Injury  
Event Description
It was reported the device pacemaker suddenly turned off during the procedure.The device was returned for repair.No patient complications were reported as a result of this event.
 
Manufacturer Narrative
This event occurred outside the us where the same model is distributed.All information provided is included in this report.Patient information is not generally available due to confidentiality concerns.(b)(4).
 
Manufacturer Narrative
Further analysis was performed.Visual inspection revealed no anomalies.Bench analysis revealed that the device powered up normally and that both battery shorting bar measurements were within specification.Conclusion: the reported event was unable to be confirmed.
 
Manufacturer Narrative
Product event summary: analysis could not confirm the reported event; the device passed all incoming functional testing.However the device log shows that the device shut down many times due to leaving the battery drawer open and the device running on the super capacitors and depleting them or due to a system error.Analysis also found the encoders were loose in the upper case.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key3994069
MDR Text Key4891417
Report Number3004593495-2014-00027
Device Sequence Number1
Product Code DTE
Combination Product (y/n)N
Reporter Country CodeMY
PMA/PMN Number
K132924
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative,company representati
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 06/24/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/08/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/26/2014
Is the Reporter a Health Professional? No
Date Manufacturer Received03/31/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-1356-2015
Patient Sequence Number1
-
-