PHYSIO-CONTROL, INC LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR; DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES) Back to Search Results
Model Number 20
Device Problem Device Inoperable (1663)
Patient Problem No Patient Involvement (2645)
Event Date 07/15/2014
Event Type  malfunction  
Manufacturer Narrative
Physio-control evaluated the device and verified the reported failure.Physio replaced the system controller and user interface pcb assemblies and observed proper device operation through functional and performance testing.The device was then returned to the customer for use.Physio further evaluated the removed pcb assemblies and determined that the cause of the reported failure was due to a failure of an integrated circuit chip, designator u61 from the system controller pcb assembly.
 
Event Description
The customer reported upon booting the device up, it would enter a constant reset cycle and would not fully power on.The device would not have the ability to be used on a patient, if needed.There was no patient use associated with the reported failure.
 
Manufacturer Narrative
(b)(4).Physio-control continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of Device
DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)
Manufacturer (Section D)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer (Section G)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer Contact
jason march
4258674000
MDR Report Key3995920
MDR Text Key15472645
Report Number3015876-2014-00918
Device Sequence Number1
Product Code LDD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063119
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 07/15/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number3202487
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/13/2014
Is the Reporter a Health Professional? Yes
Device Age9 YR
Event Location Hospital
Initial Date Manufacturer Received 08/13/2014
Initial Date FDA Received08/08/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received09/03/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/20/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-