ETHICON INC. GYNECARE TVT OBTURATOR SYSTEM; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC Back to Search Results
Catalog Number 810081L
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/16/2014
Event Type  malfunction  
Event Description
It was reported that a patient underwent an obturator sling procedure on (b)(6) 2014.During the procedure, when pulling the trocar through the tissue out the exit point, the physician noticed that the white sheath had split.This caused difficulty pulling the sheath through the tissue.The procedure was completed successfully with no adverse patient consequences.
 
Manufacturer Narrative
(b)(4).Conclusion: no conclusion can be drawn at this time.Should additional information be obtained, a supplemental 3500a form will be submitted accordingly.
 
Manufacturer Narrative
In addition, a review of the batch manufacturing records was conducted and the batch met all finished goods release criteria.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GYNECARE TVT OBTURATOR SYSTEM
Type of Device
MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 015
Manufacturer (Section G)
ETHICON SARL-NEUCHATEL
puits-godet 20 2000 neuchatel
neuchatel
SZ  
Manufacturer Contact
guillermo villa
route 22 west po box 151
somerville, NJ 08876
9082180707
MDR Report Key3998826
MDR Text Key4778520
Report Number2210968-2014-11095
Device Sequence Number1
Product Code OTN
Combination Product (y/n)N
PMA/PMN Number
K033568
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/23/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2015
Device Catalogue Number810081L
Device Lot Number3777940
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/29/2014
Initial Date FDA Received08/11/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/30/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/25/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-