AGA MEDICAL CORPORATION AMPLATZER MUSCULAR VSD OCCLUDER; CARDIAC OCCLUSION DEVICE Back to Search Results
Model Number 9-VSD-MUSC-004
Device Problem Migration or Expulsion of Device (1395)
Patient Problem Embolism (1829)
Event Date 06/17/2014
Event Type  Injury  
Event Description
A 4mm amplatzer muscular vsd occluder (muscvsd) was released from the cable then embolized from the ventricular septal defect.The muscvsd was snared and removed.No adverse patient effects were reported.
 
Manufacturer Narrative
The 4mm muscvsd was returned to sjm and decontaminated.The muscvsd was removed from the snare, grossly and microscopically examined, and no anomalies were found.The muscvsd met dimensional specifications when measured with a caliper.The muscvsd was loaded into a test loader, deployed and retracted without difficulty or deformation, under non-physiological conditions.The device history record for this product was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined the process was performed and completed in accordance with sjm specifications and procedures.There was no evidence to suggest there was an intrinsic defect in the occluder, and the cause for the embolization remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER MUSCULAR VSD OCCLUDER
Type of Device
CARDIAC OCCLUSION DEVICE
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key3999374
MDR Text Key4775376
Report Number2135147-2014-00074
Device Sequence Number1
Product Code MLV
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P040040
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 07/18/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/11/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date10/24/2017
Device Model Number9-VSD-MUSC-004
Device Lot Number1211271042
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/30/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/18/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/24/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-