The instrument was returned and evaluated.Per the customer reported complaint, failure analysis observed that the pitch up cable was found to be broken at the distal clevis hub.The cable segment that contains the crimp was still installed in clevis.The clevis did not exhibit any damage or wear marks.Additional observation was main tube damage.The distal end of the main tube exhibited various scratch marks with light material removal and a rough surface finish.The scratches were short in length and were not aligned with the tube axis.Failure analysis concluded that damage was likely due to mishandling.The endowrist® instruments instructions for use (ifu) specifically states: general precautions and warnings handle instruments with care.Avoid mechanical shock or stress that can cause damage to the instruments.The customer reported complaint does not itself constitute a reportable event; however, the broken cable and the additional finding of main tube damage with material removed, found during failure analysis could likely cause or contribute to an adverse event if to reoccur.
|
It was reported that during a da vinci surgical procedure, a split cable was found on the pk dissecting forceps instrument.There were no reports of fragments falling into a patient.The planned surgical procedure was completed and no patient harm, adverse outcome or injury was reported.
|