Data from the analyzer was received and investigated.No problems were identified based on the data logs.The event is suspected to be caused by pollution of the pco2 measurements with flush gas.This problem is very rare and is most likely due to debris in the measuring chamber such as a hair or a clot.It is not possible to identify this type of problem in calibration data, since the calibration measurements are performed on gas.This is why the calibration has not shown any irregularities.The analyzer was brought back to the local radiometer service office where the following actions were performed.The measuring chamber was replaced and furthermore pump rotor type a and type b were replaced.Gas valves have been cleaned and the membranes have been replaced electrode channel for pco2 has been checked and was ok.At last the pco2 electrode was replaced.After these activities the analyzer showed no problems during calibration and qc measurements.The analyzer will now be re-installed at the customer's site.Info regarding if this problem has been observed previously has been requested.Also a status from the customer regarding the performance of the analyzer, after a couple of weeks in function has been requested.
|