Results & conclusions: the complaint of "neurological deficit disorder" could not confirm through products testing alone.As received, the lead was cut but complete.The lead was returned cut as a result of the explant procedure.It had some discoloration on paddle and one of the terminal end lead segment.All lead segments were inspected under the microscope and no broken wires were observed.The returned ipg was functionally tested and passed all testing.Visual and functional analysis of the returned products did not reveal any anomalies that would contribute to the reported complaint.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|