Catalog Number 2B8071
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Event Description
It was reported that a vial-mate reconstitution device had particulate matter within it.After the vial (non-baxter product) was attached to the sodium chloride solution bag with the vial-mate device a dark grey particulate was observed in the solution bag.The customer stated that the particulate matter was the membrane of the vial.The customer stated that they previously used filters to prevent this from occurring but have since discontinued filter use.This occurred during use.There was no patient injury, medical intervention or adverse event associated with this event.No additional information is available.
 
Manufacturer Narrative
(b)(4).The device has been received by baxter and is currently under evaluation.Upon completion of the device evaluation, or if additional relevant information is obtained, then a follow-up mdr will be submitted.
 
Manufacturer Narrative
(b)(4).This lot was manufactured from 09/20/2013 ¿ 09/24/2013.Evaluation: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was received for evaluation.Visual inspection showed no defects.The device was connected to a solution bag and vial for functional testing with no particulate matter observed.The reported problem was not verified.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IVT DISPOSABLE
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE - CLEVELAND
911 highway 61 north
po box 1
cleveland MS 38732
Manufacturer (Section G)
BAXTER HEALTHCARE - CLEVELAND
911 highway 61 north
po box 1
cleveland MS 38732
Manufacturer Contact
christina arnt
25212 w. illinois route 120
round lake, IL 60073
2242703198
MDR Report Key4003433
MDR Text Key16086957
Report Number1416980-2014-26191
Device Sequence Number1
Product Code LHI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973654
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/18/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/12/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date09/11/2018
Device Catalogue Number2B8071
Device Lot NumberGR13I11037
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/30/2014
Is the Reporter a Health Professional? No
Date Manufacturer Received09/11/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/24/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No