Catalog Number 1001780JS-HC
Device Problems Difficult to Insert (1316); Peeled/Delaminated (1454)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/22/2014
Event Type  malfunction  
Event Description
It was reported that during insertion of the balance middleweight (bmw) guide wire into the 20/30 priority pack introducer, it was observed at the junction of the guide wire that the coating was coming off the surface of the guide wire.Resistance was felt during insertion.The guide wire was not advanced further when the damaged coating was observed.No coating entered the patient anatomy.The bmw guide wire and the 20/30 priority pack guide wire introducer were and not used.Two new devices were used to complete the case.No adverse patient effects or clinically significant delay in the procedure were reported.No additional information was provided.
 
Manufacturer Narrative
(b)(4).It is indicated that the device is not returning for evaluation; therefore a failure analysis of the complaint device could not be completed.A review of the lot history record revealed no non-conformances that would have contributed to the reported event.A query of the electronic complaint handling database revealed no other incidents for difficult to insert or peeling reported from this lot.Based on the reviewed information, no product deficiency was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACS HI-TORQUE BALANCE MIDDLEWEIGHT GUIDE WIRE WITH HYDROCOAT HYDROPHILIC COATING
Type of Device
GUIDE WIRE
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 462
Manufacturer (Section G)
BARCELONETA, PUERTO RICO REG# 3005737652
abbott vascular
26531 ynez road
temecula CA 92591 462
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key4010154
MDR Text Key20109358
Report Number2024168-2014-05184
Device Sequence Number1
Product Code DQX
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K013833
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 07/23/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/14/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2016
Device Catalogue Number1001780JS-HC
Device Lot Number4060271
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/23/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-