The customer obtained discordant but reproducible vitros total b-hcg ii results from a patient sample tested undiluted and diluted (10 fold) on a vitros 5600 integrated system.Vitros b-hcg undiluted = 1644, 1651 vs.Diluted 10x = 7983, 8495 miu/ml.Ortho clinical diagnostics conservatively considers the expected vitros total b-hcg ii results for the patient sample to be the diluted values based on similar hcg results obtained using a non-vitros method.Biased results of the magnitude and direction observed may lead to inappropriate physician action if the results were to recur undetected.The vitros total b-hcg ii results were not reported from the laboratory.There was no allegation of patient harm as a result of this event.(b)(4).
|
The investigation determined that discordant but reproducible vitros total b-hcg ii test results were obtained from a patient sample tested undiluted and diluted (10 fold) on a vitros 5600 integrated system.It is conservatively concluded that the vitros total b-hcg ii results obtained from the diluted sample are the expected hcg results based on a comparable hcg result obtained using a non-vitros method; however, this was not confirmed.There was no indication that a reagent issue contributed to the event.The root cause of this event is unknown.However, an instrument related issue or an unknown sample related event cannot be ruled out as potential contributing factors.
|