Model Number 4415
Device Problem Leak/Splash (1354)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/27/2014
Event Type  malfunction  
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the cooler heater unit was leaking internally.As a result, an alternate device was employed.The surgical procedure was completed successfully.There were no delays, no blood loss, or no adverse consequences to the patient.
 
Manufacturer Narrative
The field service representative (fsr) spoke with the perfusionist (ccp) and he thought the leak was at hansen connectors on the outside of the unit.The ccp used some grease to seal the o-ring on the hansen which fixed the reported issue.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TERUMO TCM II COOLING AND HEATING SYSTEM
Type of Device
TCM II COOLING AND HEATING SYSTEM
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEM CORP
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key4016572
MDR Text Key19591457
Report Number1828100-2014-00592
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K883603
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/27/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number4415
Device Catalogue Number4415
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/27/2014
Initial Date FDA Received07/20/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/01/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-