Model Number 20
Device Problem Device Inoperable (1663)
Patient Problem No Patient Involvement (2645)
Event Date 07/22/2014
Event Type  malfunction  
Event Description
The customer reported that their device was intermittently powering on with a solid white screen showing, which is indicative of a device lockup condition.There was no patient use associated with the reported failure.
 
Manufacturer Narrative
(b)(4).The customer advised physio-control that they have evaluated the device and verified the reported failure.The customer then replaced the system controller pcb assembly.The device was then evaluated by physio-control.Physio-control observed that the system controller pcb assembly fixed the device and observed proper device operation through functional and performance testing.The device was returned to the customer for use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of Device
DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)
Manufacturer (Section D)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer (Section G)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer Contact
jason march
4258674000
MDR Report Key4021056
MDR Text Key4876650
Report Number3015876-2014-00959
Device Sequence Number1
Product Code LDD
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K063119
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 07/22/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/19/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number3202487
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2014
Is the Reporter a Health Professional? Yes
Device Age8 YR
Event Location Hospital
Date Manufacturer Received07/22/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/30/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-