• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC., CRMD PROMOTE ACCEL CRT-D, DF-4 CONNECTOR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC., CRMD PROMOTE ACCEL CRT-D, DF-4 CONNECTOR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3215-36Q
Device Problems False Alarm (1013); Low impedance (2285)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/15/2014
Event Type  malfunction  
Event Description
New information received notes that a merlin.Net transmission and alert revealed low out of range high voltage lead impedance.The hvli trend showed a low impedance value for the lower limit, but the trend did not display any out of range measurements.Device will continue to be monitored via merlin.Net.
 
Event Description
It was reported that the patient received an alert for hvli out of range.No out of range values appeared on the visual trend.It was suspected that the alert was false due to interference during the hvli check.Programming changes were recommended.There were no adverse consequences and the patient was fine after the event.
 
Manufacturer Narrative
All information provided by manufacturer, no medwatch form was received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMOTE ACCEL CRT-D, DF-4 CONNECTOR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC., CRMD
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC., CRMD
15900 valley view court
sylmar CA 91342
Manufacturer Contact
cary lawler
645 almanor avenue
sunnyvale, CA 94085
8184932621
MDR Report Key4021833
MDR Text Key5149824
Report Number2938836-2014-14415
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2011
Device Model NumberCD3215-36Q
Device Catalogue NumberSMTFY999
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/15/2014
Initial Date FDA Received08/19/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received10/22/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/16/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age52 YR
-
-