MAUDE Adverse Event Report: PHYSIO-CONTROL, INC LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR; DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR; DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES) Back to Search Results
Model Number 20
Device Problem Device Inoperable (1663)
Patient Problem No Patient Involvement (2645)
Event Date 07/28/2014
Event Type  malfunction  
Manufacturer Narrative
Correction information: the initial medwatch report did not include the evaluation of the device performed by physio-control, which was available at the time of the submission.The device arrived for evaluation on (b)(4) 2014.Physio-control evaluated the device and verified the reported failure.Physio replaced the system controller and user interface pcb assemblies and observed proper device operation through functional and performance testing.The device was then returned to the customer for use.Supplemental information: the removed pcb assemblies were further evaluated in the failure analysis center.The cause of the reported failure was determined to be a shorted integrated circuit chip, designator u61.Pin 23 of the ic chip had a 45 ohm short to ground.
 
Event Description
The customer reported that their device would not complete its boot up cycle when powering the device on.This was discovered during testing and there was no patient use associated with the reported failure.
 
Manufacturer Narrative
(b)(4).The hospital biomed has performed an evaluation on the device and verified the reported failure.Physio-control continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of Device
DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)
Manufacturer (Section D)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer (Section G)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer Contact
jason march
4258674000
MDR Report Key4028664
MDR Text Key12374357
Report Number3015876-2014-00968
Device Sequence Number1
Product Code LDD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063119
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 07/28/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/21/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number3202487
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/21/2014
Is the Reporter a Health Professional? Yes
Device Age9 YR
Event Location Hospital
Date Manufacturer Received09/22/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/07/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-