• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH COLIBRI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH COLIBRI Back to Search Results
Catalog Number 532.001
Device Problems Melted (1385); Overheating of Device (1437)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
Customer reported device got hot and battery melted.This is report 1 of 1 for #(b)(4).
 
Manufacturer Narrative
Synthes is submitting this report as a result of remediation activities related to (b)(4) and repair files legacy review/remediation protocol-complaint handling and mdr reporting.Blank fields on this form indicate the information is unknown, unavailable or unchanged.Device listed in this report is used for treatment, not diagnosis.Any additional information received regarding this event after filing this report shall be filed on a supplemental mdr.During pre-repair assessment performed by technician, inspection revealed the device was not functioning due to motor heavy moving, jammed and seized.The device was repaired and returned to customer on 10/18/2012.The complaint is indeterminate.(b)(6).510k#: device is not distributed in the united states, but is similar to device marketed in the usa.Placeholder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COLIBRI
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
ch-4436
oberdorf
SZ 
Manufacturer (Section G)
SYNTHES GMBH
eimattstrasse 3
ch-4436
oberdorf
SZ  
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
8006207025
MDR Report Key4033830
MDR Text Key4725926
Report Number8030965-2014-00756
Device Sequence Number1
Product Code MOQ
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/10/2012
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number532.001
Device Lot Number8848
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/17/2012
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/10/2012
Initial Date FDA Received08/25/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/10/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-