• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR; DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR; DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES) Back to Search Results
Model Number 20
Device Problem Device Inoperable (1663)
Patient Problem No Patient Involvement (2645)
Event Date 05/30/2014
Event Type  malfunction  
Event Description
The customer reported that their device had been dropped but was also doing a constant reboot during the power up sequence.The customer initially reported that this reboot issue occurred before the drop but later was unable to confirm if it was before or after the drop.Upon physio-control's evaluation of the device, it was observed that the reboot was occurring due to a failure of an integrated circuit on the system controller pcb assembly.Based on the evaluation performed by physio, the reported reboot issue would have likely been occurring prior to the device being dropped.There was no patient use associated with the reported failure.
 
Manufacturer Narrative
Physio-control evaluated the device and verified the reported failure.Physio replaced the system controller and user interface pcb assemblies and then observed proper device operation through functional and performance testing.The device was then returned to the customer for use.Physio further evaluated the removed pcb assemblies and determined that the cause of the reported failure was due to a failure of an integrated circuit chip, designator u61 from the system controller pcb assembly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of Device
DC-DEFIBRILLATOR, LOW-ENERGY, (INCLUDING PADDLES)
Manufacturer (Section D)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer (Section G)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer Contact
jason march
4258674000
MDR Report Key4041090
MDR Text Key18792908
Report Number3015876-2014-00998
Device Sequence Number1
Product Code LDD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063119
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 07/16/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number3202487
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/10/2014
Is the Reporter a Health Professional? Yes
Device Age9 YR
Event Location Hospital
Initial Date Manufacturer Received 07/16/2014
Initial Date FDA Received08/27/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/27/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-