• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KERR CORPORATION SONICFILL COMPOSITE; TOOTH SHADE RESIN MATERIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KERR CORPORATION SONICFILL COMPOSITE; TOOTH SHADE RESIN MATERIAL Back to Search Results
Catalog Number 34922
Device Problem Loss of or Failure to Bond (1068)
Patient Problem No Code Available (3191)
Event Type  Injury  
Event Description
A doctor alleged that approximately seven (7) patients had experienced their restorations not fully polymerizing after placement with the sonicfill product.This is the fifth of seven (7) reports.
 
Manufacturer Narrative
Specific patient information with regard to gender, age and weight was not provided.The doctor removed the sonicfill restoration and repeated the procedure during the same visit, without further incident.To date, the patient is doing fine.A depth of cure and curing test per tip were performed on the returned device, yielding results within specifications.A dhr review revealed that there were no deviations within the manufacturing process.In addition, no similar complaints were received with regards to this lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SONICFILL COMPOSITE
Type of Device
TOOTH SHADE RESIN MATERIAL
Manufacturer (Section D)
KERR CORPORATION
1717 west collins avenue
orange CA 92867
Manufacturer (Section G)
KERR CORPORATION
1717 west collins avenue
orange CA 92867
Manufacturer Contact
kerri casino
1717 west collins avenue
orange, CA 92867
7145167634
MDR Report Key4042746
MDR Text Key4803638
Report Number2024312-2014-00569
Device Sequence Number1
Product Code EBF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K091023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Dentist
Type of Report Initial
Report Date 07/29/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/27/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Dentist
Device Expiration Date02/01/2016
Device Catalogue Number34922
Device Lot Number5113263
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/29/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/12/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
-
-