Device Problems
Device Operates Differently Than Expected (2913); Positioning Problem (3009)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/08/2014 |
Event Type
No Answer Provided
|
Event Description
|
The user facility reported that an operator tried articulating the table via the hand control during a patient procedure and the table would not respond.The operator utilized the table's manual override controls and the procedure was completed successfully.No injuries or procedural delays/cancellations were reported.
|
|
Manufacturer Narrative
|
A steris service technician inspected the table and found three out of the four table feet were out of adjustment.The technician replaced the feet, tested the table and returned it to service.The table was remanufactured in 2013 and is under steris warranty.Steris is replacing the surgical table subject of the reported event with a new table.
|
|