HILL-ROM, INC. AFFINITY 4 BED; BIRTHING BED Back to Search Results
Model Number 3700
Device Problems Collapse (1099); Component Falling (1105)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/25/2014
Event Type  malfunction  
Event Description
Hill-rom received a report from the account stating one of the casters fell off the bed resulting in the bed collapsing.The vest was located at the account in the deliver suite.There was no pt/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
 
Manufacturer Narrative
The hill-rom tech found when he arrived at the account the bed had been repaired by the hospital's service provider.A search for the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unk if the facility performs preventative maintenance on their beds.The tech performed a functional and visual check to verify the bed was functioning as designed.Based on this info, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFFINITY 4 BED
Type of Device
BIRTHING BED
Manufacturer (Section D)
HILL-ROM, INC.
batesville IN
Manufacturer Contact
john cummings
1069 state route 46 e
batesville, IN 47006
8129312869
MDR Report Key4077216
MDR Text Key18539484
Report Number1824206-2014-02146
Device Sequence Number1
Product Code KNC
Combination Product (y/n)N
PMA/PMN Number
K915779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 07/25/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/22/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number3700
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received07/25/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-