• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICRUS ENDOVASCULAR, LLC PRESIDIO 18 - CERECYTE MICROCOIL; CNV DCS COILS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRUS ENDOVASCULAR, LLC PRESIDIO 18 - CERECYTE MICROCOIL; CNV DCS COILS Back to Search Results
Catalog Number PC418083030
Device Problem Impedance Problem (2950)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/18/2010
Event Type  malfunction  
Event Description
The presidio 18 cerecyte microcoil 8 mm x 30 cm (pc418083030/f54056) won't move though the dpu wire was pushed.It was removed together with the microcatheter (details unknown).
 
Manufacturer Narrative
The presidio 18 cerecyte microcoil 8 mm x 30 cm (pc418083030/f54056) won't move though the dpu wire was pushed.It was removed together with the microcatheter (details unknown).The product was not returned for evaluation.Without the return of the device, the exact root cause of the problem reported could not be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design or quality concerns.No corrective actions will be taken at this time.This mdr is being submitted as part of a retrospective review as the result of a recent fda audit and with accordance to the requirements of code of federal regulations ¿ 21 cfr part 803, medical device reporting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESIDIO 18 - CERECYTE MICROCOIL
Type of Device
CNV DCS COILS
Manufacturer (Section D)
MICRUS ENDOVASCULAR, LLC
821 fox lane
san jose CA 95131
Manufacturer (Section G)
CODMAN AND SHURTLEFF, INC (SAN JOSE)
821 fox lane
san jose CA 95131
Manufacturer Contact
duane durbin
miami lakes, FL 33014
5088283106
MDR Report Key4085646
MDR Text Key4825824
Report Number1226348-2014-00516
Device Sequence Number1
Product Code HCG
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K002056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Health Professional
Type of Report Initial
Report Date 10/18/2010
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/19/2014
Device Catalogue NumberPC418083030
Device Lot NumberF54056
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/18/2010
Initial Date FDA Received09/12/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/01/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MICROCATHETER (DETAILS UNKNOWN)
-
-