The customer reported a leak from a coulter act diff 2 analyzer.The customer estimated the volume of the leak was a few drops and was not contained within the instrument.In addition, the instrument was generating intermittent hemoglobin (hgb) values and white blood cell (wbc) flags.The instrument operator was wearing gloves, goggles, and a lab coat at the time of the leak.There were no reports of biohazard exposure to the leak.There were no erroneous results generated and patient treatment was not impacted in connection with this event.
|
A beckman coulter (bec) field service engineer (fse) was dispatched to evaluate the instrument.The fse did not observe an active leak, but did observe hgb background failures and intermittent wbc flags.To resolve the hgb issue, the fse replaced the lamp.To resolve the flags, the fse made adjustments to the red blood cell (rbc) and wbc baths and ran clog detection calibration.Repairs were verified per established procedures.(b)(4).
|