• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS CLINITEK STATUS+; CT STATUS+

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS CLINITEK STATUS+; CT STATUS+ Back to Search Results
Catalog Number 10379675
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/18/2014
Event Type  malfunction  
Event Description
Customer reported that they received a brand new unit with broken power cord.There was no report of injury due to this event.
 
Manufacturer Narrative
The broken power cord has been received and been evaluated by siemens technical solution centre (tsc).The tsc indicated that the power supply box was appeared to be undamaged.Tsc also indicated that the power supply bosses were broken off.A new power cord will be provided to the customer.The cause for the broken power cord is unknown.
 
Manufacturer Narrative
The original equipment manufacturer (oem) of these power supplies has identified two root causes for the damaged power supply adaptors.The incorrect torque setting was selected by the production operator for the screws used to secure the power supply adaptor housings together.Mineral oil was found on the screws used to secure the housings together.The oil is used during the machining process during the production of the screws and was not removed by the screw supplier.These two root causes resulted in a combination of mechanical and chemical stress which led to the cracking of the screw pillars inside the housing top of the power supply adaptor.Siemens issued an urgent field safety notice (31983 rev.A) to notify all affected siemens customers.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINITEK STATUS+
Type of Device
CT STATUS+
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict avenue
tarrytown 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
northern road
chilton industrial estate
sudbury CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
steven andberg
2 edgewater drive
norwood, MA 02062
7812693655
MDR Report Key4096851
MDR Text Key4761275
Report Number1217157-2014-00130
Device Sequence Number1
Product Code JIL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K091216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 08/19/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/17/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number10379675
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/24/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Initial
Removal/Correction Number1217157-09-22-2014-001-R
Patient Sequence Number1
-
-