Catalog Number 26-1221
Device Problems Plunge (1462); Failure to Auto Stop (2938)
Patient Problem No Code Available (3191)
Event Date 09/10/2014
Event Type  Injury  
Event Description
Disengagement failure was noted during surgery.Perforator could not stop and caused injury to the patient's dura.It was reported that the product was not reprocessed.Bleeding was stopped and the case was completed.There was no surgical delay greater than 30 min.The patient is now followed.
 
Manufacturer Narrative
Upon completion of the investigation, a follow up report will be filed.
 
Manufacturer Narrative
Upon completion of the investigation it was noted that the reported problem of disengagement failure was not verified.The customer's perforator met functional test acceptance requirements.Proper engagement and disengagement was achieved with the perforator with every drilled hole.Device history records (perforator assembly) show all tests and inspections, including a drilling test, met specification requirements.Based on the results of this investigation no further action is required.Trends will be monitored for this and similar complaints.At the present time this complaint is closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOSABLE PERFORATOR
Type of Device
DRILLS, BURRS, TREPHINES & ACC.
Manufacturer (Section D)
CODMAN & SHURTLEFF, INC.
new bedford industrial park
new bedford MA 02745
Manufacturer Contact
james kenney
325 paramount drive
raynham, MA 02767
5088282726
MDR Report Key4101940
MDR Text Key19298421
Report Number1226348-2014-11989
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number26-1221
Device Lot NumberDG011S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/25/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/02/2014
Initial Date FDA Received09/19/2014
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/05/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/17/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-