Model Number 3186
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Bacterial Infection (1735)
Event Date 08/26/2014
Event Type  Injury  
Event Description
Device 2 of 3.Reference mfr report: 1627487-2014-02662, 02664.The pt received two leads from the lot number, and two anchors from the same lot number.It was reported the pt experienced an infection shortly after she was implanted with her scs system.The pt's daughter stated the pt was reportedly diagnosed with a staph infection, and the pt's entire scs system was removed.The pt's daughter stated the pt has reportedly recovered.
 
Manufacturer Narrative
Method: the device history and sterilization records were reviewed.Results: the device history and sterilization records reviewed found one non-conformance.The nonconforming units were replaced.The product met all final packaging/sterilization acceptance criteria prior to release for distribution.Conclusion: the cause of the reported complaint could not be determined from the review of the dhr and sterilization records.Sjm has limited info related to the pt's medical history and is unable to form an opinion as to the relevancy of the pt's history to the event reported.Sjm defers to the pt's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
plano TX
Manufacturer Contact
arnie ochoa
6901 preston rd.
plano, TX 75024
9723098090
MDR Report Key4117257
MDR Text Key15175345
Report Number1627487-2014-02663
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Type of Report Initial
Report Date 08/26/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/19/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/01/2012
Device Model Number3186
Device Lot Number3263518
Was Device Available for Evaluation? No
Date Manufacturer Received08/26/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/01/2010
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age73 YR
-
-