BECKMAN COULTER COULTER LH 750 HEMATOLOGY ANALYZER; COUNTER, DIFFERENTIAL CELL Back to Search Results
Catalog Number 6605632
Device Problems Thermal Decomposition of Device (1071); Failure to Power Up (1476)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/01/2014
Event Type  malfunction  
Event Description
The customer reported the plug to the power supply of a coulter lh 750 hematology analyzer burned causing a burning odor in the lab.The customer stated the analyzer would not power on.There was no fire or smoke; the customer did not observe sparks, arcs, or flames.The fire department was not called and a fire extinguisher was not used.There were no reports of injury and patient results were not impacted in connection with this event.
 
Manufacturer Narrative
A beckman coulter (bec) field service engineer (fse) was dispatched to evaluate the instrument.The fse determined the female side of the power cord to the power supply was melted and the male side was damaged from overheating.The fse replaced the power supply, fittings, and input power plug connection resolving the issue.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER LH 750 HEMATOLOGY ANALYZER
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s. kraemer blvd.
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th avenue
miami FL 33196
Manufacturer Contact
dung nguyen
250 s. kraemer blvd.
brea, CA 92821
7149614941
MDR Report Key4130465
MDR Text Key4941344
Report Number1061932-2014-02468
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K061574
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/01/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/30/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number6605632
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/01/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-